RIDEN RD6024-W DIY laboratory power supply 0-60V, 0-24A WiFi
Code: LA122064Free Shipping
From 100 EUR
30 Days to Return
For Registered
30 Days to Return
For Registered Customers
Free Shipping
For orders from 100 EUR
Customer Care
We handle complaints within 5 days
Quick dispatch
Within 24 hours
Related products
Product detailed description
With the Riden RD6024 laboratory power supply you can easily monitor and regulate voltage and current in various electronic circuits. This equipment is often used in laboratories, electronic project development and other applications where precise control of voltage and current is required.
The RD6024 is a useful tool for engineers, technicians and electronics enthusiasts who need a reliable and accurate voltage and current source for their applications.
Build your lab source with the Box and the Source!
Specifications:
- input voltage: 6-70V
- output voltage: 0V-60V (0.01V continuously adjustable). Max will be one volt less than the input voltage
- output current: 0-24.00A (0.001A continuously adjustable)
- typical output ripple: 250 mVh (peak-to-peak) at 6A
- output power max: 1480W
- output voltage accuracy: ±(0.5%+3 characters)
- output current accuracy: ±(0.5%+5 characters)
- efficiency: 96%
- Module dimensions: 167×81×65 mm
- Hole in panel: 160×74 mm
- weight: 730g
Attention:
- It is important that the output side is separated from the input side, otherwise the inverter may be damaged.
- If you are connecting an inductive load, such as a motor, it is important that the output current is no more than 1/3 of the rated current of the inverter. You must not exceed this limit even in the short term.
- When unpacking the inverter, care should be taken not to touch the components and board with bare hands as static electricity could damage the sensitive control and measurement electronics. Install the inverter in a workplace with anti-static equipment and wear anti-static gloves. The inverter should only be handled beyond its peripheral edges.
- For proper operation of the inverter, the input voltage should be at least 3 to 5 volts higher than the output voltage.
- If you decide to connect the inverter to a rectified voltage from a transformer or to an AC/DC converter, it is recommended to use a filter (large capacitors with a minimum capacitance of 4700 µF) at the input. If the inverter still shows instability, it is advisable to consider adding an LC filter (with a minimum 47 µH inductor and a minimum 2200 µF capacitor).
- It is not recommended to connect non-original sensors, as specifically emphasized by the manufacturer, as this can lead to damage to the inverter.
Included in delivery:
- 1pc RIDEN RD6024-W DIY laboratory power supply 0-60V, 0-24A WiFi
Remark:
- This product is not a standalone functional unit and may require professional installation.
- Product images are for illustration purposes only and may sometimes differ from the actual appearance of the item. However, this does not change its basic properties.
- When working with high current, it is necessary to ensure heat dissipation.
Be the first who will post an article to this item!
Add a rating
F
Krabička
František
Dobrý den, je k tomu prosím i kovová krabička, kam vložím zdroj i jeho napájení?
Děkuji
Reply
Reply
MF
Krabička
Martin Frajdl
Dobrý den, tato krabička není součástí balení.